Instruments Care
COLORED HALO STAINS WITH SHARP CONTOURS
Problem: Deposits or
Organic Substance.
Possible Cause: Unknown.
Solution: Rinse
instrument thoroughly and immediately after use in distilled or de-mineralized
water.
COLORED HALO STAINS AND THE SHARP CONTOURS ARE NOT PRESENT
Problem: Metal
Deposits.
Possible Cause: In flushing water
or steam there are iron, copper, manganese etc: are present.
Solution: Use
distilled or de-mineralized water.
WHITISH STAINS
Problem: Phosphate
Deposits.
Possible Cause: Unknown.
Solution: Check
Composition of Cleaning Solution.
YELLOW TO DARK BROWN COLORS
Problem: Residue
Deposits.
Possible Cause: Old cleaning
solutions / Incorrectly cleaned instruments.
Solution: Change baths
after each use/ Rinse instruments immediately after use in distilled or
de-mineralized water.
BLACK STAINS
Problem: Caustic
Reaction.
Possible Cause: Detergent with
low pH content (6 or less).
Solution: Use
Neutral Detergent.
BLUISH-BLACKISH STAINS
Problem: Metallic
Coating or Platting.
Possible Cause: Metals of
different types have been sterilized together such as: for instance, stainless
steel with Silver / Chrome / etc.
Solution: Sterilize
only the similar metal instruments together.
SEVERAL COLORS
Problem: Excess Heat
(Chromatic Oxide).
Possible Cause: The autoclave is
not operating properly or has been set to very high temperature (This can cause
metal to lose some of its properties such as strength and hardness, etc.).
Solution: Correct the autoclave temperature.